【Scope of Application】
Used to test whether the resistance values of medical and daily protective masks during use meet the standards for exhalation and inhalation of the human body.
[Related Standards]
GB2626-2019, GB/T32610-2016, TAJ1001-2015, etc.
【Technical Specifications】
1. Flowmeter, range: (0 to ±100) L/min, accuracy: 3%
2. Micro-pressure sensor, range: ±1000 Pa, accuracy: 1 Pa
3. Air supply device, air supply volume approximately 200 L/min, maximum exhaust pressure 200 kPa, maximum negative pressure 90 kPa
4. Default air supply volume is (85 ± 1) L/min (can be set as needed)
5. Test head mold specification can be replaced as needed (standard model: GB2626)
6. Qualified indicators: Exhalation resistance less than 250 Pa, inhalation resistance less than 350 Pa
7. Power supply: AC 220V ± 10% 50Hz 500W
8. Instrument size: 620 × 650 × 850 mm (L × W × H)
9. Weight: approximately 30 kg